Enhanced chromatographic determination of nicotine in human plasma: Applied to human volunteers
Document Type
Article
Publication Date
12-15-2015
Abstract
© 2015 Bassam M. Ayoub et al. Development of enhanced UPLC-UV method for determination of nicotine in human plasma was achieved on a Symmetry® C18 column (100 mm × 2.1 mm, 2.2 μm) applying isocratic elution based on Methanol: Acetonitrile: Phosphate Buffer (pH: 2.7) with the ratio (20:30:50, v/v/v) as a mobile phase. The ultraviolet detector was operated at 260 nm. The mobile phase was pumped through the column at a flow rate of 0.2 mL min-1. The column temperature was adjusted to 50ºC and the injection volume was 2 μL. Quinine was selected as an internal standard (IS) due to its structure similarity to nicotine having basic pyridine ring to optimize the liquid liquid extraction procedure using diethyl ether coupled with vacuum evaporation at 40ºC. Validation parameters for nicotine were found to be acceptable over the concentration range of 2.5-50 ng ml-1. The application of the proposed method on four healthy human volunteers was approved by the ethical committee. The study was carried out under fasting conditions and the concerned subjects were informed about the objectives and possible risks involved in the study. The proposed method proved to be simple and fast which is a major advantage to analyze large number of samples per day using the accelerated vacuum evaporation technique. The method showed satisfactory data for all the parameters tested within the limits for bioanalytical assays. The lower limit of quantification (LLOQ) permits the application of the method for further pharmacological and clinical studies.
Recommended Citation
Ayoub, B.M., Mohamady, S., Hendy, M.S., Elmazar, M.M. Enhanced chromatographic determination of nicotine in human plasma: Applied to human volunteers (2015) International Journal of Biomedical Science, 11 (4), pp. 185-189.