Abstract

Determination of a widely used antihypertensive combination of atenolol and hydrochlorothiazide was achieved by rapid and eco-friendly high-performance liquid chromatography method combined with fluorescence detection. The response surface methodology is conducive to the complete separation of the two drugs in a shorter analysis time. The separation of the mixture was achieved using an Inertsil C18 analytical column (150 × 4.6 mm, 5 µ). The mobile phase used was ethanol: potassium dihydrogen phosphate at pH 3 (65:35 v/v) and the flow rate was 0.7 mL/min. The fluorescence detector operated at excitation and emission wavelengths of 230 and 310 nm (atenolol) and 270 and 375 nm (hydrochlorothiazide). The linearity of the developed method covered a concentration of atenolol of 0.05–5 µg/mL and a concentration of hydrochlorothiazide of 0.02–1 µg/mL. The greenness of the developed method was evaluated by analytical eco-scale and the recently reported evaluation method, that is, green analytical procedure index, and it was found to be an excellent, sensitive, and green alternative to the reported methods. The developed method was validated according to the ICH guidelines and compared with the reference method. No significant difference was found in terms of accuracy